TP 17 650 S
SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a broad spectrum of industries with diverse applications.
Dry-block function
The optimum thermal coupling from the block to test item is achieved with the correct adapter sleeve. Ideally, the internal diameter of the sleeve is 0.5 mm larger than the outer diameter of the test item. With the aid of the adapter sleeve, straight temperature sensors of virtually any length and diameter can be calibrated. The dry block covers the entire temperature range without the need to change the calibration medium. Viscosity, flash points or outgassing are of no concern.
Applications
Assembly and commissioning – Manufacturing and production – Maintenance and service – Quality assurance and test equipment monitoring – Repair
TP 17650 S |
* At an ambient temperature of 20 °C |
Dry block | |
Temperature range* | RT…650 °C |
Tolerance | ±0,4 °C |
Stability | ±0,05 °C |
Resolution | 0,01 °C in the range of RT…99,99 °C, else 0,1 °C |
Block | Ø 28 mm / depth 150 mm |
- Overview
-
TP 17 650 S
SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a broad spectrum of industries with diverse applications.
Dry-block function
The optimum thermal coupling from the block to test item is achieved with the correct adapter sleeve. Ideally, the internal diameter of the sleeve is 0.5 mm larger than the outer diameter of the test item. With the aid of the adapter sleeve, straight temperature sensors of virtually any length and diameter can be calibrated. The dry block covers the entire temperature range without the need to change the calibration medium. Viscosity, flash points or outgassing are of no concern.Applications
Assembly and commissioning – Manufacturing and production – Maintenance and service – Quality assurance and test equipment monitoring – Repair
- Technical Specifications
-
TP 17650 S
* At an ambient temperature of 20 °C
Dry block Temperature range* RT…650 °C Tolerance ±0,4 °C Stability ±0,05 °C Resolution 0,01 °C in the range of RT…99,99 °C, else 0,1 °C Block Ø 28 mm / depth 150 mm - Downloads