MPRF High Temperature
Adaptable to almost any situation, the MPRF Temperature data logger’s small size increases versatility and precision by allowing you to position the data logger very near or inside your target products.
- Wireless Temperature Data Loggers
- Real-Time Data
- FDA 21 CFR Part 11 compliant software
Applications
Cold Chain – Pharmaceuticals – Produce/Meat/Dairy – Agriculture & Floral – Office & Warehouse – Museums – Chemicals – Laboratories – Server room
MPRF High Temperature |
|
Operating Temperature:
|
HiTemp***: +100°C (+212°F) to +360°C (+680°F) w/Thermal Pack
HiTemp EXT***: -20°C (-40°F) to +400°C (+752°F) w/Thermal Pack
|
Temperature Accuracy: |
-80°C to -40°C (-112°F to -40°F):±0.3°C (±0.5°F)
-40°C to +140°C (-40°F to +284°F):±0.1°C (±0.2°F)
|
Resolution: |
0.001°C
|
Operating Pressure: |
Up to 10 BAR
|
Wireless Range: |
100 feet (30 meters) Line of Sight
|
Radio Frequency: | 2.4 GHz ISM Band |
Size:
|
Width: 0.91 in (23.2mm)
Length: 3.7 in (94mm)**
|
Weight: | 2.3 oz (66 grams)** |
Material: | stainless steel and ULTEM |
Probe Diameter: |
Rigid probe diameter: 0.125”/3.2mm or .188” (4.76mm) available
Bendable probe diameter: 0.079”/2mm
|
Temperature: | PT1000 Class A RTD |
Calibration: | Factory calibrated (NIST traceable) or Field Calibration |
Battery Type: | AA Lithium (field-replaceable) |
Battery Life: | 6 months to 1 year (depending on usage) |
Battery Weight:
|
Standard Temperature Battery: 0.6 oz (17 grams)
Low Temperature Battery: 0.67 oz (19 grams)
|
Clock accuracy: | 10 seconds per 24 hours (25°C) |
Certifications: |
ATEX (for intrinsic safety)
FCC Class B digital device (part 15 compliance)
|
Memory |
8,192 time/temperature observations or 16,384 observations in compressed data mode
|
Sample rate
|
Operator programmable
• 5 sec. to 1 hr interval with radio
• 1 sec. to 1 hr interval without radio
|
**Body only.
***Thermal Packs required above +140°C
- Overview
-
MPRF High Temperature
Adaptable to almost any situation, the MPRF Temperature data logger’s small size increases versatility and precision by allowing you to position the data logger very near or inside your target products.
- Wireless Temperature Data Loggers
- Real-Time Data
- FDA 21 CFR Part 11 compliant software
Applications
Cold Chain – Pharmaceuticals – Produce/Meat/Dairy – Agriculture & Floral – Office & Warehouse – Museums – Chemicals – Laboratories – Server room
- Technical Specifications
-
MPRF High Temperature
Operating Temperature:HiTemp***: +100°C (+212°F) to +360°C (+680°F) w/Thermal PackHiTemp EXT***: -20°C (-40°F) to +400°C (+752°F) w/Thermal PackTemperature Accuracy: -80°C to -40°C (-112°F to -40°F):±0.3°C (±0.5°F)-40°C to +140°C (-40°F to +284°F):±0.1°C (±0.2°F)Resolution: 0.001°COperating Pressure: Up to 10 BARWireless Range: 100 feet (30 meters) Line of SightRadio Frequency: 2.4 GHz ISM Band Size:Width: 0.91 in (23.2mm)Length: 3.7 in (94mm)**Weight: 2.3 oz (66 grams)** Material: stainless steel and ULTEM Probe Diameter: Rigid probe diameter: 0.125”/3.2mm or .188” (4.76mm) availableBendable probe diameter: 0.079”/2mmTemperature: PT1000 Class A RTD Calibration: Factory calibrated (NIST traceable) or Field Calibration Battery Type: AA Lithium (field-replaceable) Battery Life: 6 months to 1 year (depending on usage) Battery Weight: Standard Temperature Battery: 0.6 oz (17 grams)Low Temperature Battery: 0.67 oz (19 grams)Clock accuracy: 10 seconds per 24 hours (25°C) Certifications: ATEX (for intrinsic safety)FCC Class B digital device (part 15 compliance)Memory 8,192 time/temperature observations or 16,384 observations in compressed data modeSample rate Operator programmable• 5 sec. to 1 hr interval with radio• 1 sec. to 1 hr interval without radio**Body only.***Thermal Packs required above +140°C - Downloads