MPIII Dual Temp

Monitor 2 Data Points with 1 Logger

The DataTrace Dual Probe Temperature Logger allows the user to measure 2 data points with a single logger. With 2 probes, this data logger is ideal for either collecting data from within a product and from the ambient conditions within a process or for collecting data from 2 separate areas of the product in question. This functionality provides the user with great flexibility as well as a great deal of information that a single probe logger cannot offer.

From chilling, freezing and room temperature to pasteurization, EtO sterilization and depyrogenation, MPIII Temp, HiTemp and HiTemp EXT Loggers provide accurate data over a wide range with remarkable ease.

Their small size increases versatility and precision by allowing you to position Loggers very near or inside monitored items.

 

  • Wireless Temperature Data Logger
  • Measures Ambient and Product Simultaneously
  • 2 Data Points in Single Logger
  • Field calibration/battery change
  • 16,000 NIST-traceable data points
  • FDA 21 CFR Part 11 Compliant
  • Intrinsically Safe

MPIII Dual Temp

 

Size: Width: 0.7 in. (18mm) Length: 0.9 in. (23mm)**
Weight: 0.5 oz. (15 grams)**
Material:
316 stainless steel; Sapphire IR windows
Memory: 16,000 time/temperature observations
Sample rates:
Operator programmable 1 sec. to 1 hr. intervals
Operating Temperatures:
MPIII Temp and MPIII Dual Temp: -20°C (-4°F) to +140°C (+284°F)
MPIII HiTemp*: +100°C (+212°F) to +360°C (+680°F) w/Thermal Pack
MPIII HiTemp EXT*: -20°C (-4°F) to +400°C (+752°F) w/Thermal Pack
Resolution:
0.001°C
Temperature Accuracy:
MPIII Temp/Dual: ±0.1°C (±.2°F)
MPIII HiTemp*: ± 0.5°C (±0.9°F)
MPIII HiTemp EXT*: ± 0.25°C (±0.45°F)
<200°C (392°F)± 0.5°C (± 0.9°F)
≥200°C (392°F)
Operating Pressure: For all 3 models: Up to 10 Bar
Calibration:
Factory calibrated (NIST traceable)
Field calibration (MPIII Temp/Dual only)
Battery Type: Lithium (field-replaceable)
Clock accuracy: 26 seconds per 24 hours
Certifications: ATEX (for intrinsic safety)
Overview

MPIII Dual Temp

Monitor 2 Data Points with 1 Logger

The DataTrace Dual Probe Temperature Logger allows the user to measure 2 data points with a single logger. With 2 probes, this data logger is ideal for either collecting data from within a product and from the ambient conditions within a process or for collecting data from 2 separate areas of the product in question. This functionality provides the user with great flexibility as well as a great deal of information that a single probe logger cannot offer.

From chilling, freezing and room temperature to pasteurization, EtO sterilization and depyrogenation, MPIII Temp, HiTemp and HiTemp EXT Loggers provide accurate data over a wide range with remarkable ease.

Their small size increases versatility and precision by allowing you to position Loggers very near or inside monitored items.

 

  • Wireless Temperature Data Logger
  • Measures Ambient and Product Simultaneously
  • 2 Data Points in Single Logger
  • Field calibration/battery change
  • 16,000 NIST-traceable data points
  • FDA 21 CFR Part 11 Compliant
  • Intrinsically Safe
Technical Specifications

MPIII Dual Temp

 

Size: Width: 0.7 in. (18mm) Length: 0.9 in. (23mm)**
Weight: 0.5 oz. (15 grams)**
Material:
316 stainless steel; Sapphire IR windows
Memory: 16,000 time/temperature observations
Sample rates:
Operator programmable 1 sec. to 1 hr. intervals
Operating Temperatures:
MPIII Temp and MPIII Dual Temp: -20°C (-4°F) to +140°C (+284°F)
MPIII HiTemp*: +100°C (+212°F) to +360°C (+680°F) w/Thermal Pack
MPIII HiTemp EXT*: -20°C (-4°F) to +400°C (+752°F) w/Thermal Pack
Resolution:
0.001°C
Temperature Accuracy:
MPIII Temp/Dual: ±0.1°C (±.2°F)
MPIII HiTemp*: ± 0.5°C (±0.9°F)
MPIII HiTemp EXT*: ± 0.25°C (±0.45°F)
<200°C (392°F)± 0.5°C (± 0.9°F)
≥200°C (392°F)
Operating Pressure: For all 3 models: Up to 10 Bar
Calibration:
Factory calibrated (NIST traceable)
Field calibration (MPIII Temp/Dual only)
Battery Type: Lithium (field-replaceable)
Clock accuracy: 26 seconds per 24 hours
Certifications: ATEX (for intrinsic safety)
Downloads