The HygroGen2 satisfies the stringent demands of quality and compliance regulations with its ability to calibrate instruments over their full working range, and is embraced throughout the pharmaceutical industry as the leading instrument in its class and is favored in ISO17025 humidity calibration laboratories across the world. Being portable and robust, it can be set up at the installation location allowing for complete system qualification.

The HygroGen2 works like a “mobile calibration laboratory” and is aimed at companies that regularly have to calibrate a large number of sensors. The calibrator enables simple, flexible calibration with the advantage that the devices to be calibrated can be quickly integrated back into the operating processes.

The rock solid stability, minimal thermal gradients, and rapid time for set-point changes of HygroGen2-XL are now replicated in a new family member: HygroGen2-S, which is more handy due to its lower weight.

 

Features
• Generates stable humidity and temperature conditions (<±0.1 %RH, <±0.01 °C)
• Calibration solution for the laboratory and on-site
• Humidity equilibrium typically in only 15 minutes
• Calibrate multiple probes simultaneously
• ≪AutoCal≫ automated calibration (optional)
• ≪Remote API≫ logging and control with third party software (optional)*
• Integrated FDA 21 CFR part 11 compliant Rotronic HW4 software

Control range max. 5…95 %RH and 0…60 °C
(2…99 %RH / -5…60 °C with range extension)
Control stability <±0.01 °C,<± 0.1 %RH
Temperature gradient <±0.05 °C (15…50 °C), <±0.1 °C (5…60 °C), <±0.15 °C (0…5 °C)
Humidity gradient 0.1 %RH or better
Typical time to set-point <15 minutes (5…95 %RH)
Temperature sensor PT100 1/3 DIN Class B
Humidity sensor Hygromer
Control probe accuracy ±0.8 %RH (23 °C ±5) / ±2 %RH (0…60 °C) /
±0.1 k (23 °C ±5) / ±0.3 k (0…60 °C)
Desiccant Molecular sieve
Saturator Front panel fill, on-screen level indication
Water quality Distilled or deionised water
Water volume approx. 80 ml
Chamber volume 20 litres, effective working volume 17 litres
External interface USB (9 ports), Ethernet
Power consumption 110…230 VAC, 50…60 Hz, 3 A
External connections 6mm temperature controlled sample loop connections
Enclosure material Powder coated aluminium
Dimensions 800 x 620 x 410 mm
Weight 37 kg
Overview

HygroGen2 XL

 

The HygroGen2 satisfies the stringent demands of quality and compliance regulations with its ability to calibrate instruments over their full working range, and is embraced throughout the pharmaceutical industry as the leading instrument in its class and is favored in ISO17025 humidity calibration laboratories across the world. Being portable and robust, it can be set up at the installation location allowing for complete system qualification.

The HygroGen2 works like a “mobile calibration laboratory” and is aimed at companies that regularly have to calibrate a large number of sensors. The calibrator enables simple, flexible calibration with the advantage that the devices to be calibrated can be quickly integrated back into the operating processes.

The rock solid stability, minimal thermal gradients, and rapid time for set-point changes of HygroGen2-XL are now replicated in a new family member: HygroGen2-S, which is more handy due to its lower weight.

 

Features
• Generates stable humidity and temperature conditions (<±0.1 %RH, <±0.01 °C)
• Calibration solution for the laboratory and on-site
• Humidity equilibrium typically in only 15 minutes
• Calibrate multiple probes simultaneously
• ≪AutoCal≫ automated calibration (optional)
• ≪Remote API≫ logging and control with third party software (optional)*
• Integrated FDA 21 CFR part 11 compliant Rotronic HW4 software

Technical specification
Control range max. 5…95 %RH and 0…60 °C
(2…99 %RH / -5…60 °C with range extension)
Control stability <±0.01 °C,<± 0.1 %RH
Temperature gradient <±0.05 °C (15…50 °C), <±0.1 °C (5…60 °C), <±0.15 °C (0…5 °C)
Humidity gradient 0.1 %RH or better
Typical time to set-point <15 minutes (5…95 %RH)
Temperature sensor PT100 1/3 DIN Class B
Humidity sensor Hygromer
Control probe accuracy ±0.8 %RH (23 °C ±5) / ±2 %RH (0…60 °C) /
±0.1 k (23 °C ±5) / ±0.3 k (0…60 °C)
Desiccant Molecular sieve
Saturator Front panel fill, on-screen level indication
Water quality Distilled or deionised water
Water volume approx. 80 ml
Chamber volume 20 litres, effective working volume 17 litres
External interface USB (9 ports), Ethernet
Power consumption 110…230 VAC, 50…60 Hz, 3 A
External connections 6mm temperature controlled sample loop connections
Enclosure material Powder coated aluminium
Dimensions 800 x 620 x 410 mm
Weight 37 kg